OWON SDS220S

OWON SDS220S-200MHz Desktop Oscilloscope User Manual

Model: SDS220S

1. Bubuka

The OWON SDS220S-200MHz is a high-performance dual-channel desktop oscilloscope integrated with a 25MHz waveform generator. This device features a 200MHz bandwidth, 1GSa/s sample rate, and a 20M record length, making it suitable for various electronic measurement and analysis tasks. It supports 30 automatic waveform measurements, FFT Spectrum Analysis, and PC communication via USB for remote control and secondary development using SCPI and LabVIEW. This manual provides essential information for the safe and effective operation of your SDS220S-200MHz oscilloscope.

OWON SDS220S-200MHz Multifunctional Digital Oscilloscope

Figure 1.1: OWON SDS220S-200MHz Multifunctional Digital Oscilloscope overview, highlighting its 2 channels, 20M record length, 1GSa/s real-time sample rate, and 25MHz waveform generator.

2. Émbaran Kasalametan

To prevent electric shock or personal injury, and to avoid damage to the oscilloscope or connected equipment, observe the following safety precautions:

  • Grounding anu leres: Ensure the oscilloscope is properly grounded. The instrument is grounded through the power cord's grounding conductor. To avoid electric shock, the grounding conductor must be connected to earth ground.
  • Connect and Disconnect Properly: Connect the probe output to the oscilloscope input before connecting the probe to the circuit under test. Disconnect the probe from the circuit under test before disconnecting it from the oscilloscope.
  • Perhatikeun Sadaya Peunteun Terminal: To prevent fire or shock hazard, observe all ratings and markings on the instrument. Consult the product manual for more information about ratings before making connections to the instrument.
  • Use Proper Probes: Use only probes supplied with the instrument or recommended by the manufacturer. Ensure probes are rated for the voltage and current being measured. The included P4250 High Voltage Probe has a 100:1 attenuation ratio and is rated for 2KV Working Voltage (Vp-p).
  • Avoid Exposed Circuitry: Do not touch exposed connections or components when power is present.
  • Ulah beroperasi kalawan disangka Gagal: If you suspect damage to the instrument, have it inspected by qualified service personnel.
  • Ngajaga kaayaan garing: Do not operate the instrument in wet or damp kaayaan.
  • Operate in a Well-Ventilated Area: Pastikeun ventilasi ditangtoskeun pikeun nyegah overheating.

3. Produk Langkungview

3.1 Fitur konci

  • 200MHz Bandwidth, Dual Channel
  • 1GSa/s Sample Rate (Dual-channel Independent)
  • 20M Record Length
  • Built-in 25MHz Waveform Generator (1CH AWG, 125MSa/s Real-time Sample Rate, 8K Arbitrary waveform length)
  • 7-inch TFT LCD (800*480 pixels resolution)
  • Support for 30 Automatic Waveform Measurements
  • Analisis Séktrum FFT
  • SCPI and LabVIEW support for remote control and secondary development
  • USB Device interface for PC communication
  • 16 Waveform Storage locations
  • Counter Frékuénsi
  • Compact and lightweight design (1.1KG / 2.42 lbs)

3.2 Komponén Panel hareup

OWON SDS220S Front Panel Layout

Figure 3.1: Front panel of the SDS220S, showing the display, button and knob control area, power switch, function keys, probe compensation output, oscilloscope signal input ports (CH1, CH2), and signal generator output port.

3.3 Komponén Panel pungkur

OWON SDS220S Rear Panel Layout

Figure 3.2: Rear panel of the SDS220S, illustrating the foldable handle, ventilation holes, power socket, USB Device interface for PC connection, and the integrated stand.

3.4 Included Accessories: P4250 High Voltage Ulikan

The SDS220S includes an extra P4250 High Voltage Probe. This probe is designed for high voltage measurements and features:

  • Bandwidth: 250MHz
  • Attenuation Ratio: 100:1
  • Résistansi input: 100MΩ
  • Max Input Voltage: 2KV Working Voltage (Vp-p)
P4250 High Voltage Probe and Specifications

Figure 3.3: The P4250 High Voltage Probe, an included accessory, with its detailed specifications table.

4. Setélan

4.1 Ngabongkar sareng Inspeksi

  1. Carefully remove the oscilloscope and all accessories from the packaging.
  2. Inspect the device for any signs of damage during transit. If damage is found, contact your supplier immediately.
  3. Pariksa heula sadaya barang anu aya dina daptar barang bawaan.

4.2 Sambungan Daya

  1. Connect the provided power cord to the power socket on the rear panel of the oscilloscope.
  2. Plug the other end of the power cord into a grounded AC power outlet. Ensure the power supply voltage cocog sareng sarat anu ditangtukeun dina labél alat.

4.3 Mimiti Power-On

  1. Press the power switch located on the front panel to turn on the oscilloscope.
  2. The device will perform a self-test and display the startup screen.

4.4 Sambungan Probe sareng Kompensasi

  1. Connect the BNC connector of the oscilloscope probe to one of the input channels (CH1 or CH2) on the front panel.
  2. Attach the probe tip to the probe compensation output (5V/1kHz signal) on the front panel.
  3. Adjust the compensation trimmer on the probe until a flat-top square wave is displayed on the screen. This ensures accurate measurements.

5. Parentah Operasi

5.1 Basic Oscilloscope Operation

The SDS220S allows for precise visualization and analysis of electrical signals.

Oscilloscope measuring a signal

Figure 5.1: The SDS220S oscilloscope in operation, displaying a waveform on its screen while connected to a circuit.

  1. Connecting to a Circuit: Connect the compensated probe to the desired test point on your circuit.
  2. Skala Vertikal (Volt/Div): Ngagunakeun éta Nangtung knob to adjust the vertical sensitivity, changing the voltage scale per division on the display.
  3. Posisi nangtung: Ngagunakeun éta Posisi knob in the Vertical section to move the waveform up or down on the screen.
  4. Skala Horisontal (Waktos/Div): Ngagunakeun éta Horizontal knob to adjust the time base, changing the time scale per division.
  5. Posisi Horisontal: Ngagunakeun éta Posisi knob in the Horizontal section to shift the waveform left or right.
  6. Dipicuna: The Pemicu section controls when the oscilloscope starts acquiring data. Adjust the trigger level to stabilize the waveform. The SDS220S supports Edge and Video trigger types.
  7. Pangukuran otomatis: Pencét éta Ukur button to access the automatic measurement menu. The SDS220S supports 30 types of automatic measurements (e.g., Period, Frequency, Vpp, Vmax, Vmin, Rise Time, Fall Time).
30 Automatic Waveform Measurements

Figure 5.2: The SDS220S displaying the menu for 30 automatic waveform measurements, allowing for detailed signal analysis.

5.2 Waveform Generator Operation

The built-in 25MHz waveform generator can produce various standard and arbitrary waveforms.

Oscilloscope and Waveform Generator in use

Figure 5.3: The SDS220S functioning as both an oscilloscope (left) and a waveform generator (right), demonstrating its dual capabilities.

  1. Kaluaran Sambungkeun: Connect the signal generator output port to the desired input of your circuit or to an oscilloscope channel for monitoring.
  2. Select Waveform: Use the function keys to select the desired waveform type (Sine, Square, Ramp, Pulse, Arbitrary). The SDS220S offers 5 standard waveforms and over 160 kinds of arbitrary waveforms.
  3. Saluyukeun Parameter: Use the control knobs and menu options to set the frequency (up to 25MHz), amplitude (5mVpp to 3Vpp), and DC offset of the generated waveform.

5.3 Fitur canggih

  • FFT Spectrum Analysis: Pencét éta Matematika button and select FFT to analyze the frequency components of a signal.
  • Mode XY: This mode is used to display the relationship between two signals, often for phase comparison.
  • Save/Recall: The oscilloscope allows saving and recalling up to 16 waveforms for later analysis or comparison.
  • Konektipitas PC: Connect the oscilloscope to a PC via the USB Device interface for remote control and data transfer. The device supports SCPI commands and LabVIEW for custom applications.
Remote Control and Secondary Development via USB

Figure 5.4: The SDS220S connected to a laptop via USB, demonstrating its capability for remote control and secondary development using SCPI and LabVIEW software.

6. Pangropéa

6.1 Ngabersihan Alat

To maintain the performance and appearance of your oscilloscope:

  • Pegatkeun sambungan kabel listrik saméméh beberesih.
  • Make lawon anu lemes dampened with a mild detergent solution to clean the exterior surfaces. Do not use abrasive cleaners or solvents.
  • For the display, use a soft, lint-free cloth specifically designed for screens.
  • Ensure no liquid enters the device through ventilation holes or ports.

6.2 Kaperawatan Probe

  • Store probes carefully to prevent damage to the tips or cables.
  • Regularly check probe cables for cuts or damage. Replace damaged probes immediately.
  • Recalibrate probes periodically, especially after significant temperature changes or if measurements appear inaccurate.

7. Cara ngungkulan

Bagian ieu nyayogikeun solusi pikeun masalah umum anu anjeun tiasa patepang.

MasalahMungkin Cukang lantarananaSolusi
Taya kakuatanPower cord not connected; Power outlet faulty; Device fuse blownPariksa sambungan kabel listrik; Tés stop kontak; Hubungi layanan pikeun ngaganti sekering.
Teu aya bentuk gelombang anu ditampilkeunProbe not connected; Input signal too small/large; Trigger settings incorrect; Channel turned offEnsure probe is connected to input and circuit; Adjust Vertical Scale (Volts/Div); Adjust trigger level and mode; Ensure channel is enabled.
Bentuk gelombang distortedProbe compensation incorrect; Probe bandwidth insufficient; Signal integrity issuesPerform probe compensation; Use a probe with appropriate bandwidth (e.g., P4250 for high frequency); Check signal source.
Waveform unstableTrigger level or mode incorrect; Noise in signalAdjust trigger level and mode; Use averaging mode to reduce noise.
Waveform generator output not workingOutput cable not connected; Output enabled setting incorrect; Parameters set to zeroConnect output cable; Ensure waveform generator output is enabled in menu; Check frequency and ampsetélan litude.

8. Spésifikasi

8.1 Parameter osiloskop

ParameterSpésifikasi
Bandwidth Analog200MHz
Saluran2
Sampling Laju1GSa/s
Max Record Length20M
Masukan GandengDC, AC, Tanah
Rentang Sensitivitas Vertikal2 mV/div ~ 10 V/div
Resolusi nangtung8 bits (2 Channels Simultaneously)
Waveform Storage16 Bentuk gelombang
Jenis pemicuEdge, Video
Waktos Range Base2ns/div ~ 1000s/div, Step by 1-2-5
Waveform Math+, -, ×, ÷, FFT
Input impedansi1MΩ±2%, In Parallel with 20 pF±5 pF
Modeu ScanNormal, Peakdetect, Averaging
Antarbeungeut KomunikasiUSB Host jeung Alat

8.2 Waveform Generator Parameters

ParameterSpésifikasi
Max Frequency Output25MHz
Saluran1CH AWG
Sample Laju125MSa / s
Arbitrary Waveform Length8K
Kaluaran Amplituit5mVpp to 3 Vpp (50Ω)
Resolusi nangtung14 bit
Waveforms bakuSinus, Square, Ramp Wave, Pulse Wave, Arbitrary Wave
Arbitrary WaveformsSinc, Exponential Rise, Exponential Decline, Gaussian, more than 160 kinds
DC Offset±(3V) (High Resistance)

8.3 Spésifikasi Umum

ParameterSpésifikasi
tampilan7-inch TFT LCD (800*480 pixels)
Dimensi produk11.85 x 2.76 x 6 inci (301 x 70 x 152 mm)
Item Beurat2.42 pon (1.1 KG)
Suhu Operasi-10°C ~ 55°C
Kalembaban Operasi≤90% (41°C) to ≤60% (50°C)
Luhurna Operasi3000m
Detailed Oscilloscope and Waveform Generator Specifications

Figure 8.1: A comprehensive table detailing the specifications for both the oscilloscope and waveform generator functions of the SDS220S.

OWON SDS220S Dimensions

Figure 8.2: The physical dimensions of the OWON SDS220S, highlighting its compact and ultra-thin body design.

9. Garansi jeung Rojongan

9.1 Émbaran jaminan

OWON products are manufactured under strict quality standards. This product is covered by a limited warranty against defects in materials and workmanship. The specific warranty period and terms may vary by region and retailer. Please retain your proof of purchase for warranty claims.

9.2 Rojongan Téknis

For technical assistance, troubleshooting, or service inquiries, please contact your local OWON distributor or refer to the official OWON website for support resources and contact information. When contacting support, please have your product model (SDS220S) and serial number available.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.