1. Bubuka
This manual provides comprehensive instructions for the safe and efficient operation of your Voltcraft VC 630-2 Analog Oscilloscope. Designed for professional use in fields such as assistance, laboratory work, education, and hobby electronics, this 2-channel oscilloscope offers a bandwidth of 0 (DC) to 30 MHz. Please read this manual thoroughly before using the device to ensure proper functionality and to prevent damage or injury.

Figure 1: Voltcraft Brand Logo
2. Parentah Kasalametan
Always adhere to the following safety guidelines to prevent electric shock, fire, or damage to the device.
- Sasayogian tanaga: Ensure the oscilloscope is connected to a power source matching the specified voltage (115 V/AC or 230 V/AC). Use only the provided power cord.
- Lingkungan: Operate the device in a dry, well-ventilated area. Avoid exposure to moisture, extreme temperatures, or direct sunlight.
- Ventilasi: Ulah ngahalangan liang ventilasi. Aliran hawa anu cekap penting pisan pikeun nyegah panas teuing.
- Grounding: The oscilloscope must be properly grounded. Do not defeat the grounding plug.
- Masukan Voltage: The device is rated for CAT II 300 V. Do not apply voltages exceeding this rating to any input terminal.
- Palayanan: Refer all servicing to qualified service personnel. Do not attempt to open or repair the device yourself.
- beberesih: Cabut daya sateuacan dibersihkeun. Anggo lawon anu lemes sareng garing. Entong nganggo pembersih cair atanapi aerosol.
3. Produk Langkungview
The Voltcraft VC 630-2 is a robust analog oscilloscope featuring two input channels and a clear display for waveform analysis. Below is an image illustrating the main components of the device.

Figure 2: Front Panel of the Voltcraft VC 630-2 Oscilloscope. This image shows the display screen, various control knobs for vertical and horizontal adjustments, trigger settings, and input connectors for Channel 1 and Channel 2.
Fitur konci:
- 2-Channel Input
- 30 MHz Bandwidth
- High-Sensitivity Triggering
- Input éksternal
- X-Y Operation Mode
- Z-Input
- Phase Meter Functionality
- 50 Ω Output for Frequency Counters
4. Setélan
4.1 Ngabongkar sareng Inspeksi
- Carefully remove the oscilloscope from its packaging.
- Inspect the device for any signs of physical damage that may have occurred during transit.
- Verify that all included accessories, such as the power cord and user manual, are present.
4.2 Sambungan Daya
- Ensure the power switch on the oscilloscope is in the OFF position.
- Connect the provided power cord to the power input on the rear panel of the oscilloscope.
- Plug the other end of the power cord into a grounded electrical outlet that supplies either 115 V/AC or 230 V/AC, as appropriate for your region.
4.3 Mimiti Power-On
- After connecting the power, switch the oscilloscope ON.
- Observe the display for initial power-up sequence and self-test indicators.
- Allow a few moments for the device to stabilize before proceeding with measurements.
5. Parentah Operasi
5.1 Kadali Dasar
- Saklar kakuatan: Located on the front panel, used to turn the device ON/OFF.
- Intensity/Focus: Adjusts the brightness and sharpness of the trace on the screen.
- Position (Vertical/Horizontal): Moves the waveform vertically or horizontally on the display.
5.2 Vertical Deflection (VOLTS/DIV)
The VOLTS/DIV knob controls the vertical sensitivity of the input signal. It ranges from 5 mV/Div to 5 V/Div. Select the appropriate setting based on the amplitude of the signal you are measuring.
- Input Coupling (AC/DC/GND):
- AC: Blocks the DC component of the signal, allowing only the AC component to be displayed.
- DC: Displays both AC and DC components of the signal.
- GND: Disconnects the input signal and grounds the input, useful for setting the zero reference level.
- CH1/CH2 Inputs: Connect your probes to the BNC connectors for Channel 1 and Channel 2.
5.3 Horizontal Sweep (TIME/DIV)
The TIME/DIV knob controls the horizontal sweep speed, determining how much time each horizontal division represents. Adjust this to view the desired number of cycles of your waveform.
5.4 Nyimbulkeun
Triggering stabilizes the waveform on the display. The VC 630-2 features high-sensitivity triggering.
- tingkat: Saluyukeun voltage level at which the trigger occurs.
- lamping: Selects whether the trigger occurs on the rising or falling edge of the signal.
- Sumber: Selects the trigger source (e.g., CH1, CH2, External).
5.5 Fungsi Maju
- X-Y Operation: Allows the oscilloscope to display one input signal against another, useful for Lissajous figures and phase measurements.
- Z-Input: Modulates the intensity of the trace, often used for blanking or brightening specific parts of the waveform.
- Phase Meter: Utilize this function for precise phase difference measurements between two signals.
- 50 Ω Output: An additional 50 Ω output is provided for connecting external frequency counters, enhancing the versatility of the oscilloscope.
6. Pangropéa
6.1 beberesih
To maintain the performance and appearance of your oscilloscope:
- Salawasna pegatkeun sambungan kabel listrik sateuacan diberesihan.
- Paké lawon lemes, garing, lint bébas ngusap surfaces exterior.
- Keur kokotor bandel, rada dampen the cloth with water or a mild, non-abrasive cleaner. Do not allow any liquids to enter the device.
- Avoid using harsh chemicals, solvents, or abrasive cleaning agents.
6.2 Panyimpenan
When not in use, store the oscilloscope in a cool, dry place, away from direct sunlight, dust, and extreme temperatures. Protect it from physical shocks.
6.3 Kalibrasi
The Voltcraft VC 630-2 is factory calibrated. For continued accuracy, periodic professional calibration may be required depending on usage and application. Please note that the product description indicates "factory (without certificate)".
7. Cara ngungkulan
Bagian ieu nyayogikeun solusi pikeun masalah umum anu anjeun tiasa patepang.
| Masalah | Mungkin Cukang lantaranana | Solusi |
|---|---|---|
| Teu aya tampilan/Pareuman | Kabel listrik teu nyambung; Stop kontak rusak; Saklar listrik pareum. | Check power cord connection; Test power outlet; Ensure power switch is ON. |
| No trace on screen | Intensity too low; Focus out of adjustment; Vertical/Horizontal position off-screen; Input coupling set to GND. | Adjust Intensity and Focus knobs; Center Position controls; Change Input Coupling from GND to AC or DC. |
| Bentuk gelombang anu teu stabil | Trigger level incorrect; Trigger source incorrect; Signal too noisy. | Adjust Trigger Level; Select correct Trigger Source (CH1, CH2, EXT); Check signal integrity. |
| Bentuk gelombang distorted | Input signal too high (clipping); Probe compensation incorrect; Bandwidth exceeded. | Adjust VOLTS/DIV to a higher setting; Compensate probe; Ensure signal frequency is within 30 MHz bandwidth. |
8. Spésifikasi
Technical specifications for the Voltcraft VC 630-2 Analog Oscilloscope (Model: GOS-630).
| Parameter | Nilai |
|---|---|
| Nomer modél | GOS-630 |
| Bandwidth | 0 (DC) to 30 MHz |
| Saluran | 2 |
| Masukan Gandeng | AC, DC, GND |
| Vertical Deflection | 5 mV - 5 V / Div |
| Diménsi Monitor | 10 x 8 cm |
| Sasayogian tanaga | 115 V/AC, 230 V/AC |
| Beurat | 7.8 kg |
| Overvoltage Kategori | CAT II 300V |
| Produsén | VOLTCRAFT |
9. Jaminan & Dukungan
Voltcraft products are manufactured to high quality standards and come with a standard warranty. For specific warranty terms and conditions, please refer to the documentation provided with your purchase or visit the official Voltcraft websitus.
For technical support, troubleshooting assistance beyond this manual, or service inquiries, please contact Voltcraft customer service. Contact details can typically be found on the manufacturer’s websitus atanapi dina bungkusan produk Anjeun.