Extech PH110

Buku Panduan Meter pH ExStik Extech PH110 Tahan Air

Brand: Extech | Model: PH110

1. Bubuka

The Extech PH110 Waterproof ExStik pH Meter is designed for accurate pH measurements in various substances. This device features a refillable pH electrode, allowing for extended use and consistent accuracy by replenishing the reference solution. Its robust, waterproof design ensures reliability in diverse environments.

2. Fitur konci

  • Refillable pH electrode for extended service life.
  • Analog bargraph display on LCD for trend indication.
  • Waterproof construction to IP57 standard.
  • Automatic 1, 2, or 3 point calibration with buffer solution recognition.
  • Flat Surface Electrode capable of measuring pH in liquids, semi-solids, and solids.

3. Eusi Paket

The Extech PH110 Waterproof ExStik pH Meter package includes:

  • Extech PH110 pH Meter with refillable pH electrode
  • 1.5mL bottle of KCI (Potassium Chloride) solution
  • Two reference junctions
  • Two electrode removal tools
  • Cap pelindung
  • Sample cangkir
  • Four 3V CR-2032 batteries (pre-installed or included separately)
  • 48-inch (1.2m) neck strap

4. Setélan

4.1 Pamasangan batré

  1. Ensure the meter is off.
  2. Panggihan kompartemen batré.
  3. Insert four 3V CR-2032 Lithium Metal batteries, observing correct polarity.
  4. Tutup kompartemen batré anu aman.

4.2 Electrode Preparation

  1. Cabut tutup pelindung tina éléktroda.
  2. Rinse the electrode tip with distilled water.
  3. If the electrode appears dry, soak it in electrode storage solution or pH 7 buffer for at least 30 minutes before first use.

5. Operasi

5.1 Kalibrasi

Calibration is essential for accurate measurements. The PH110 supports 1, 2, or 3 point calibration and automatically recognizes buffer solutions.

  1. Turn on the meter by pressing the ON/OFF kancing.
  2. Prepare fresh pH buffer solutions (e.g., pH 4, 7, and 10). These are sold separately.
  3. Pencét jeung tahan tombol KAL button until 'CAL' appears on the display.
  4. Immerse the electrode into the first buffer solution (e.g., pH 7). The meter will automatically recognize the buffer and calibrate. Wait for the reading to stabilize.
  5. For 2 or 3 point calibration, rinse the electrode with distilled water and immerse it into the next buffer solution (e.g., pH 4, then pH 10). The meter will continue to calibrate automatically.
  6. Ensure the display shows 'CAL' during the calibration process.

5.2 Pangukuran

The flat surface electrode allows for versatile measurements.

  1. After calibration, rinse the electrode with distilled water.
  2. Celupkeun éléktroda kana sample to be measured. For liquids, ensure the electrode tip is fully submerged. For semi-solids or solids, ensure good contact with the flat surface electrode.
  3. Allow the reading to stabilize on the display. The analog bargraph will indicate trends in acidity or alkalinity.
  4. To switch between pH and temperature display modes, press the MODÉ kancing.
  5. After measurement, rinse the electrode thoroughly with distilled water and replace the protective cap.
Extech PH110 Waterproof ExStik pH Meter displaying a pH reading.

Gambar 1: The Extech PH110 pH meter, a handheld device with a digital display showing pH and temperature readings. It features three control buttons: MODE, CAL, and ON/OFF.

6. Pangropéa

6.1 Electrode Refilling

The refillable electrode design allows for prolonged accuracy and lifespan.

  1. When the reference solution depletes or accuracy declines, use the provided removal tools to access the reference junctions.
  2. Carefully refill the electrode with the 1.5mL KCI solution.
  3. Ensure the reference junctions are properly re-seated after refilling.
Hands refilling the Extech PH110 pH meter electrode with filling solution.

Gambar 2: A nutup-up view of hands refilling the Extech PH110 pH meter's electrode module with a specialized electrode filling solution from a small bottle. This process ensures the meter's continued accuracy and extends electrode life.

6.2 Beberesih sareng Panyimpenan

  • Always rinse the electrode with distilled water after each use to prevent contamination.
  • The flat surface electrode is designed for easy cleaning. Gently wipe with a soft cloth if necessary.
  • Store the electrode with its protective cap filled with electrode storage solution (or pH 7 buffer if storage solution is unavailable) to keep the sensor hydrated. Do not store dry.

7. Cara ngungkulan

If the meter exhibits inaccurate readings or operational issues, consider the following:

  • Bacaan teu akurat: Recalibrate the meter using fresh buffer solutions. Ensure the electrode is clean and properly hydrated. Check if the reference solution in the electrode needs refilling.
  • Henteu aya masalah tampilan/daya: Verify that the batteries are correctly installed and have sufficient charge. Replace batteries if necessary.
  • Réspon Leungit: The electrode might be dry or contaminated. Clean and rehydrate the electrode, then recalibrate.

8. Spésifikasi

SpésifikasiRincian
ProdusénExtech
Nomer BagianPH110
Item Beurat3.87 ons
Dimensi produk6.8 x 1.6 x 1.4 inci
Nomer Modél ItemPH110
Batré4 Lithium Metal batteries (CR-2032) required (included)
Sumber DayaBatré Powered
Peunteun waterproofstandar IP57
Kalibrasi1, 2, or 3 point automatic buffer recognition
Tipe éléktrodaRefillable Flat Surface Electrode

9. Émbaran jaminan

Extech provides a full one-year warranty on parts and workmanship for the PH110 pH Meter, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modifications.

10. Rojongan

For further assistance, technical support, or to purchase accessories and replacement parts, please refer to the official Extech Instruments websitus atanapi ngahubungi departemén layanan palangganna. Inpormasi kontak biasana tiasa dipendakan dina bungkusan produsén atanapi websitus.

Dokumén Patali - PH110

Praview Extech ExStik™ Waterproof pH Meters PH100 & PH110 User Guide
Comprehensive user guide for the Extech ExStik™ Waterproof pH Meters, models PH100 and PH110. Learn about operation, calibration, maintenance, and specifications.
Praview Buku Panduan Pangguna Extech ExStik Alat Ukur pH Tahan Air PH100 ​​& PH110
Comprehensive user manual for the Extech ExStik Waterproof pH Meters, models PH100 and PH110. Provides detailed instructions on operation, calibration, electrode maintenance, battery replacement, safety precautions, specifications, and troubleshooting for these high-accuracy pH testing devices.
Praview Extech ExStik II EC400 Conductivity, TDS, Salinity, Temperature Meter User Manual
Comprehensive user manual for the Extech ExStik II EC400 meter, detailing its features for measuring conductivity, total dissolved solids (TDS), salinity, and temperature. Includes setup, operation, calibration, maintenance, troubleshooting, and specifications.
Praview Extech Comprehensive Catalog: Environmental Meters, Anemometers, and Test Equipment
Explore Extech's extensive range of high-quality environmental meters, thermo-anemometers, gas detectors, water quality testers, sound meters, light meters, and other precision test and measurement instruments designed for industrial, HVAC/R, and professional applications.
Praview Extech PH220 Waterproof Palm pH Meter User Manual
User manual for the Extech PH220 Waterproof Palm pH Meter. Covers introduction, meter description, operation, calibration, measurement, storage, battery replacement, maintenance, troubleshooting, and specifications.
Praview EXTECH 45170 4-in-1 Kalembaban, Suhu, Aliran Udara, sareng Manual Pamaké Méter Cahaya
Manual pamaké pikeun EXTECH 45170, méteran 4-in-1 pikeun ngukur kalembaban, suhu, aliran hawa, sareng cahaya. Ngawengku operasi, spésifikasi, sareng inpormasi garansi.