1. Bubuka
The Hanna Instruments HI9147-04 is a water-resistant dissolved oxygen (DO) meter designed for aquaculture applications. It provides accurate measurements of dissolved oxygen in water, crucial for monitoring water quality and supporting aquatic life.
Key features include automatic temperature compensation, manual salinity and altitude compensation, a backlit LCD screen for clear readings, and a galvanic DO probe with a 4-meter cable that requires no conditioning time, allowing for quick measurements.

Image shows the Hanna Instruments HI9147-04 dissolved oxygen meter. The main unit is black with a digital display showing '10.0 mg/L', '60.8 °F', and '9.7 g/L Salinity'. Below the display are six circular buttons: On/Off, CAL, %, BL ▲, SAL, mg/L, and °C/°F ▼. A galvanic probe with a 4-meter cable is connected to the right side of the meter.
2. Naon dina Box
Upon unpacking your HI9147-04 Dissolved Oxygen Meter, please verify that all items are present and in good condition:
- HI9147-04 Dissolved Oxygen Meter
- HI 76409/4 probe with 4m (13’) cable
- (2) Spare membranes
- HI 7042S electrolyte solution (30 mL bottle)
- (3) aki-aki 1.5V AAA
- HI 710016 shockproof rubber boot
- Obeng
- Instruksi Manual
3. Setélan
3.1 Pamasangan batré
The meter operates on three 1.5V AAA batteries. To install or replace batteries:
- Teang kompartemen batré dina tonggong méter.
- Buka panutup kompartemen batré.
- Insert the three 1.5V AAA batteries, ensuring correct polarity as indicated inside the compartment.
- Tutup panutup kompartemen batréna.
3.2 Sambungan Panyilidikan
Connect the galvanic DO probe to the designated port on the top right side of the meter. Ensure a secure connection to prevent water ingress and ensure proper electrical contact.
3.3 Kalibrasi Awal
Before first use and periodically, calibrate the meter for accurate readings. Calibration should be performed in air.
- Daya dina méteran.
- Pencét éta KAL button. The '%' indicator on the display will blink.
- With the probe exposed to air (not immersed in water), turn the trimmer (usually a small screw on the meter) until the display reads 100%.
- Pencét éta KAL button again to confirm the calibration. The meter is now calibrated.
4. Operasi
4.1 Powering on / Pareum
Pencét éta Hurung/Pareuman button located on the meter's keypad to power the device on or off.
4.2 Nyokot Pangukuran
To take a dissolved oxygen measurement:
- Ensure the meter is calibrated.
- Neuleumkeun usik kana sample water. Ensure the membrane is fully submerged and there are no air bubbles trapped around it.
- Allow sufficient time for the reading on the display to stabilize before recording the measurement.
4.3 Compensation Settings
The HI9147-04 offers compensation features for accurate readings in various conditions:
- Kompensasi salinitas: Pencét éta SAL button to manually adjust for salinity. Refer to your sample's known salinity for accurate readings.
- Pampasan Altitude: The meter features manual altitude compensation. Adjust this setting based on your operating altitude for precise measurements. Consult the full manual for detailed instructions on setting altitude.
- Santunan Suhu: The meter automatically compensates for temperature variations, ensuring accurate DO readings across different temperatures.
4.4 Pilihan Unit
You can switch between different measurement units:
- Pencét éta mg/L button to toggle between dissolved oxygen concentration in mg/L (parts per million, ppm) and % saturation.
- Pencét éta °C/°F ▼ button to toggle between Celsius and Fahrenheit temperature units.
5. Pangropéa
5.1 Kaperawatan Probe
To ensure the longevity and accuracy of your DO probe:
- Keep the probe membrane clean and free from debris.
- Rinse the probe with distilled or deionized water after each use.
- Store the probe properly, typically with the protective cap filled with electrolyte solution, to keep the membrane hydrated.
5.2 Membrane Replacement
The probe membrane should be replaced periodically or if it becomes damaged, scratched, or fouled. Refer to the included spare membranes and the detailed instructions provided with the probe for proper replacement procedures.
5.3 Electrolyte Solution Refill
Ensure the probe is filled with the HI 7042S electrolyte solution. The electrolyte solution needs to be replaced when the membrane is changed or if it becomes contaminated. Refill as needed, following the specific instructions for the probe.
5.4 Cleaning the Meter
Wipe the meter's exterior with a damp cloth. Do not use abrasive cleaners, solvents, or immerse the meter in water, as it is water-resistant but not fully submersible.
6. Cara ngungkulan
If you encounter issues with your HI9147-04 meter, refer to the following common problems and solutions:
- Bacaan anu teu akurat atanapi teu stabil:
- Ensure the probe is clean and properly calibrated.
- Check the electrolyte solution level and membrane condition; replace if necessary.
- Verify that salinity and altitude compensation settings are correct for your sample.
- Allow sufficient time for the probe to stabilize in the sample.
- Difficulty Calibrating:
- Confirm the probe is in air during calibration.
- Ensure the trimmer is adjusted correctly to reach 100%.
- If issues persist, check the probe membrane and electrolyte solution for contamination or depletion.
- Méter Teu Nyala:
- Pariksa pamasangan batré sareng pastikeun batréna henteu béak. Ganti batré upami diperlukeun.
- Pastikeun kompartemen batré ditutup aman.
- Damaged Cable:
- Inspect the probe cable for any visible damage, kinks, or breaks.
- A damaged cable may require probe replacement to restore functionality.
7. Spésifikasi
| Parameter | Nilai |
|---|---|
| Rentang | 0.0 to 50.0 mg/L (ppm) O2, 0 to 600% O2, -5.0 to 50.0 °C |
| Resolusi | 0.1 mg/L or 1% (O2) / 0.1 °C |
| Akurasi | ±1% of reading (O2) / ±0.2 °C |
| Sumber Daya | Three 1.5V AAA batteries (included) |
| Diménsi | 164 x 76 x 45mm / 6.5 x 3.0 x 1.8 inches (H x W x D) |
| Weight (with HI 76409/4 probe) | 511g / 1.12 pon |
| Nomer Modél Item | HI9147-04 |
| ASIN | B0085X2GZ6 |
| Produsén | Hanna Instrumén |
8. Émbaran jaminan
This Hanna Instruments product is covered by a manufacturer's warranty against defects in materials and workmanship. For specific warranty terms and conditions, including duration and coverage details, please refer to the warranty documentation included with your product or visit the official Hanna Instruments websitus. Simpen bukti pameseran anjeun pikeun klaim garansi.
9. Rojongan sarta Kontak Émbaran
For further assistance, technical support, or to purchase accessories and replacement parts, please visit the official Hanna Instruments websitus atanapi ngahubungan departemen layanan palanggan maranéhna:
Hanna Instruments Store: Visit Hanna Instruments Store on Amazon
For direct support, please refer to the contact information provided in your product's packaging or on the main Hanna Instruments corporate websitus.