This manual provides comprehensive instructions for the setup, operation, maintenance, and troubleshooting of the Oakton Instruments WD-35419-10 Series pH 700 Benchtop Meter. It is designed to assist users in achieving accurate and reliable pH measurements in laboratory and industrial settings. Please read this manual thoroughly before operating the instrument.
Eusi pakét
Pastikeun yén sadaya barang aya sareng henteu ruksak nalika ngabongkar:
- pH 700 Benchtop Meter unit
- Double-Junction Glass pH Electrode
- Automatic Temperature Compensation (ATC) Probe
- Electrode Stand
- Power Adapter (110/220 VAC compatible)
- Manual Pitunjuk (dokumén ieu)
Produk Leuwihview

Figure 1: Oakton Instruments WD-35419-10 Series pH 700 Benchtop Meter. The image displays the meter's digital display unit, a double-junction glass pH electrode immersed in a green liquid in a beaker, and an automatic temperature compensation (ATC) probe, all supported by an adjustable stand.
The Oakton Instruments pH 700 Benchtop Meter is designed for precise pH and temperature measurements. Its compact footprint makes it suitable for various laboratory environments. Key features include a large LCD display, intuitive button interface, and compatibility with a double-junction pH electrode for enhanced accuracy and longevity.
Disetél
- Ngabongkar komponén: Sacara saksama miceun sadaya komponén tina bungkusan.
- Posisi Stand: Place the electrode stand on a stable, level surface next to the meter.
- Connect the Electrode: Connect the double-junction glass pH electrode to the BNC connector on the back of the meter. Ensure a secure connection.
- Connect the ATC Probe: Plug the ATC probe into its designated port on the meter.
- Sambungkeun Kakuatan: Plug the power adapter into the meter's power input, then connect it to a 110/220 VAC power outlet.
- Daya mimiti Hurung: Press the power button to turn on the meter. The display should illuminate.
Parentah Operasi
Kalibrasi
Regular calibration is essential for accurate pH measurements. Calibrate the meter using at least two standard pH buffer solutions (e.g., pH 4.01, 7.00, and 10.00).
- Nyiapkeun Buffer: Ensure buffer solutions are at room temperature.
- Éléktroda Bilas: Rinse the pH electrode and ATC probe with distilled water and gently blot dry.
- Mimitian Calibration: Refer to the meter's on-screen prompts or specific button sequence (usually a "CAL" button) to enter calibration mode.
- Buffer Kahiji: Immerse the electrode and ATC probe into the first buffer solution (e.g., pH 7.00). Wait for the reading to stabilize, then confirm.
- Second Buffer: Rinse the electrode and probe, then immerse them into the second buffer solution (e.g., pH 4.01 or 10.00). Wait for stabilization and confirm.
- Kalibrasi lengkep: The meter will indicate when calibration is complete.
Nyokot Pangukuran
- Éléktroda Bilas: Rinse the calibrated electrode and ATC probe with distilled water.
- Immerse in Sample: Immerse the electrode and ATC probe into the sample solution. Ensure the electrode junction and temperature sensor are fully submerged.
- Ngastabilkeun Maca: Allow sufficient time for the reading to stabilize. The meter will typically display a "READY" indicator when stable.
- Record Reading: Note the pH and temperature values displayed.
- Beresih Saatos Paké: Rinse the electrode and probe thoroughly with distilled water after each measurement.
Pangropéa
- Panyimpenan éléktroda: Always store the pH electrode in its protective cap filled with electrode storage solution or pH 7 buffer. Never store in distilled water.
- beberesih: Clean the electrode regularly to prevent buildup. For general cleaning, soak in a mild detergent solution, then rinse thoroughly. For stubborn deposits, consult the electrode's specific cleaning instructions.
- Meter Cleaning: Wipe the meter's exterior with a damp lawon. Ulah make bahan kimia kasar atawa cleaners abrasive.
- Frékuénsi Calibration: Calibrate the meter daily or before each use for optimal accuracy, especially when measuring samples with significantly different pH values.
Pamérésan masalah
| Masalah | Mungkin Cukang lantaranana | Solusi |
|---|---|---|
| Bacaan teu stabil | Dirty electrode; air bubbles; electrode not fully immersed; expired buffer; damaged electrode. | Clean electrode; gently tap electrode to remove bubbles; ensure full immersion; use fresh buffer; replace electrode. |
| Bacaan teu akurat | Improper calibration; expired buffer; electrode not conditioned; temperature difference. | Recalibrate with fresh buffers; condition electrode; ensure ATC probe is functioning. |
| Méter teu hurung | Power adapter disconnected; faulty power outlet; meter malfunction. | Check power connections; try a different outlet; contact technical support. |
spésifikasi
- Jumlah modél: WD-35419-10
- Ukuran produk: 12.3 x 9.2 x 5 inci
- beurat: 4 pon
- Syarat Daya: 110 / 220 VAC
- Jenis éléktroda: Double-Junction Glass pH Electrode
- Kompensasi Suhu: Automatic (with ATC Probe)
- Pabrikan: Thermo Scientific (as listed on product page)
- Negara asal: Cina
- Tapak suku: Compact, over 40% smaller than comparable benchtop meters.
Garansi sareng Rojongan
For warranty information, technical support, or service inquiries, please refer to the documentation provided with your purchase or contact Oakton Instruments customer service directly. Keep your purchase receipt as proof of purchase.