HACH 2100Q01

Hach 2100Q Portable Turbidimeter Instruction Manual

Modél: 2100Q01

Bubuka

The Hach 2100Q Portable Turbidimeter is designed for accurate and reliable turbidity measurements in various environments, including field testing, plant monitoring, and water quality analysis for regulatory compliance. This manual provides essential information for the proper setup, operation, and maintenance of your 2100Q Turbidimeter.

Hach 2100Q Portabel Turbidimeter

Image: The Hach 2100Q Portable Turbidimeter, a compact device with a digital display showing a turbidity reading of 0.10 NTU.

Fitur konci:

  • EPA Compliant Design: Meets USEPA Method 180.1 criteria for regulatory reporting.
  • Ukur akurat: Measures turbidity across a 0-1000 NTU range.
  • Guided Calibration: On-screen assisted calibration, including Rapidcal™ option.
  • Data logging: Stores up to 500 measurements with detailed information.
  • Portabel sareng awét: Rugged design suitable for field and plant operations.

Disetél

1. Unpacking jeung Inspection

Carefully unpack all components from the carrying case. Verify that all items listed in the packing slip are present and undamaged. Report any missing or damaged items to Hach customer service immediately.

2. Pamasangan batré

The Hach 2100Q requires four (4) AA batteries for operation.

  1. Locate the battery compartment cover on the back of the instrument.
  2. Open the cover by sliding or unlatching it according to the indicator.
  3. Pasangkeun opat batré AA, pastikeun polaritasna (+/-) bener sakumaha anu dituduhkeun di jero kompartemenna.
  4. Tutup panutup kompartemen batréna.

3. Daya Awal Hurung

Press the power button (usually indicated by a circle with a vertical line) to turn on the turbidimeter. The display should illuminate, and the device will perform a self-check.

Parentah Operasi

1. Kalibrasi

Regular calibration ensures accurate turbidity measurements. The 2100Q features on-screen guided calibration.

  1. Turn on the turbidimeter.
  2. Navigate to the "Calibrate" or "Calibration" menu option using the arrow keys.
  3. Follow the on-screen prompts to insert the required calibration standards (e.g., 20 NTU, 100 NTU, 800 NTU).
  4. For quick low-level calibration, utilize the Rapidcal™ option if available.
  5. Press "Read" or "Done" as prompted after each standard is inserted and measured.
  6. Ensure the calibration is successfully completed and saved.
Hach 2100Q Turbidimeter Calibration Screen

Image: Close-up of the Hach 2100Q screen during calibration, showing "Cal: StablCal® 99.4 NTU" and options for different NTU standards.

Inserting a sample vial into Hach 2100Q Turbidimeter

Image: A hand inserting a sample vial into the measurement cell of the Hach 2100Q Turbidimeter for calibration or measurement.

2. Nyandak Pangukuran

The 2100Q measures turbidity in Nephelometric Turbidity Units (NTU).

  1. Ensure the turbidimeter is calibrated.
  2. Nyiapkeun s Anjeunample according to standard laboratory practices, ensuring the sample cell is clean and free of air bubbles or scratches.
  3. Selapkeun sample cell into the measurement compartment, aligning any marks as instructed by the device or manual.
  4. Close the lid of the measurement compartment.
  5. Press the "Read" button. The device will display the turbidity reading in NTU.
Hach 2100Q Turbidimeter held in hand showing a reading

Image: A hand holding the Hach 2100Q Turbidimeter, displaying a turbidity reading of 103 NTU on its screen.

3. Data Logging and Transfer

The 2100Q automatically stores up to 500 measurements. Each entry includes a timestamp, operator ID, and calibration status.

  • Viewdata: Access the "Data Log" or "History" menu option to review ukuran disimpen.
  • Mindahkeun Data: For transferring data to a computer, an optional USB+Power module is available. Connect the module to the turbidimeter and your computer, then follow the software instructions for data export.

Pangropéa

1. Ngabersihan Instrumen

Keep the exterior of the turbidimeter clean using a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the measurement cell compartment is free of dust and debris.

2. Sample Cell Care

Beresih sample cells thoroughly before each use. Fingerprints, scratches, or residue on the cell can affect measurement accuracy. Use a lint-free cloth to wipe the exterior of the cell.

3. Ngagantian batré

Replace batteries when the low battery indicator appears on the display. Refer to the "Battery Installation" section under Setup for instructions.

4. Optical System Care

The 2100Q uses a two-detector optical system. Avoid touching the optical components inside the measurement compartment. If cleaning is necessary, consult the full Hach 2100Q user manual for specific instructions to prevent damage.

Pamérésan masalah

This section addresses common issues you might encounter with your Hach 2100Q Turbidimeter.

MasalahMungkin Cukang lantarananaSolusi
Alatna teu bisa dihurungkeun.Batré paéh atawa salah dipasang.Ganti batré, mastikeun polaritasna bener.
Bacaan anu teu akurat atanapi henteu konsisten.
  • Dirty or scratched sampsél.
  • Alat kaluar tina kalibrasi.
  • Air bubbles in sample.
  • Beresih sample cell thoroughly.
  • Perform a full calibration.
  • Degas sample if necessary; tap cell gently to remove bubbles.
Pesen "Error" dina layar.Specific error code will indicate the issue.Refer to the comprehensive Hach 2100Q user manual for specific error code interpretations and solutions.

If you encounter issues not listed here or if the suggested solutions do not resolve the problem, please contact Hach Technical Support.

spésifikasi

Below are the technical specifications for the Hach 2100Q Portable Turbidimeter (Model 2100Q01).

FiturRincian
Nomer modél2100Q01
Rentang pangukuran0 - 1000 NTU
minuhanMeets USEPA Method 180.1 criteria
Sumber CahayaTungsten Filament Lamp
Sumber Daya4 aki aki diperyogikeun
Kamampuh Data LoggingNepi ka 500 pangukuran
Sistim optikTwo-detector system for color and stray light compensation
Ukuran (Kira-kira.)16 x 10 x 8 inci (Pakét Dimensi)
Beurat (Kira-kira.)4 Pon (Beurat Paket)

Garansi sareng Rojongan

Émbaran jaminan

Specific warranty terms and conditions for the Hach 2100Q Portable Turbidimeter are provided with the product documentation at the time of purchase. Please refer to the warranty card or the full user manual for detailed information regarding coverage, duration, and claims procedures. For the most current warranty details, contact Hach Company directly.

Bantosan Téknis

For technical assistance, troubleshooting guidance, or service inquiries, please contact Hach Company customer support.

  • Pabrikan: PERUSAHAAN HACH
  • Websitus: www.hach.com (Please visit the official Hach website for contact information, support resources, and additional documentation.)

Sumberdaya tambahan

No official product videos from the seller were found in the provided data. For further operational guidance or visual demonstrations, please refer to the official Hach Company websitus atawa ngahubungan saluran rojongan maranéhanana.

Technician performing water quality testing with a portable device

Image: A technician in a hard hat and safety glasses using a portable water quality testing device in an industrial setting, demonstrating typical field use.

Dokumén Patali - 2100Q01

Praview Manual Pangguna Turbidimeter Portabel Hach 2100Q sareng 2100Qis
Manual pangguna ieu nyayogikeun pitunjuk anu lengkep pikeun ngoperasikeun sareng ngajaga turbidimeter portabel Hach 2100Q sareng 2100Qis. Diajar ngeunaan spésifikasi, pamasangan, antarmuka pangguna, operasi canggih, sareng ngungkulan masalah.
Praview Buku Panduan Dasar HACH TL2360 - Panduan Pangukuran Kekeruhan
Buku pituduh pikeun turbidimeter laboratorium HACH TL2360. Diajar ngeunaan pamasangan, operasi, kalibrasi, sareng pangropéa pikeun pangukuran turbiditas cai anu akurat.
Praview Hach Model 2100N Laboratory Turbidimeter Instruction Manual
This instruction manual provides comprehensive guidance for the Hach Model 2100N Laboratory Turbidimeter, covering its operation, specifications, measurement procedures, calibration, maintenance, and troubleshooting for accurate turbidity analysis in laboratory environments.
Praview Hach Methods Quick Reference Guide: Chemical Analysis Instruments and Reagents
A comprehensive guide to Hach's chemical analysis methods, detailing test parameters, EPA approvals, instrument compatibility, and product numbers for water quality testing.
Praview HACH Conductivity Meter: USEPA Direct Measurement Method 8160 Guide
This guide details the HACH conductivity measurement method (USEPA 0.01 µS/cm to 200.00 mS/cm, Method 8160) for water and wastewater analysis. It covers instrument setup, test preparation, unit conversions, interferences, accuracy checks, and summary of the method.
Praview HACH FH950 Portable Flow Meter Basic User Manual
This document is the Basic User Manual for the HACH FH950 portable flow meter. It provides essential information for operating, installing, and maintaining the device for accurate water flow and depth measurements in various environments.