Apera Instruments AI311

Apera Instruments AI311 PH60 Waterproof pH Pocket Tester Kit User Manual

Brand: Apera Instruments

Modél: AI311

1. Bubuka

The Apera Instruments PH60 Pocket pH Tester is designed for reliable and easy pH measurement in general water solutions. This includes applications such as hydroponics, horticulture, aquaculture, pools & spas, water treatment, cooling towers, environmental monitoring, beverage making, and educational settings. The meter features a replaceable pH probe (PH60-E) specifically for general water solutions. It is also compatible with three other types of pH electrodes and an ORP probe (sold separately) for specialized applications like solid samples and surface tests.

Catetan: A few drops of 3M KCL liquid (included in the kit) should be added to low ion-strength test samples (e.g., distilled water or RO water) to significantly accelerate stabilization without affecting the pH reading.

Close-up of the PH60 pH sensor bulb

Gambar: Deukeut view of the high-quality glass pH sensor bulb, designed for accuracy and stability.

Infographic showing various applications for the PH60 pH tester

Image: Visual representation of diverse applications including hydroponics, pools, brewing, drinking water, water treatment, aquariums, and aquaculture.

2. Naon dina Box

Your Apera Instruments PH60 Waterproof pH Pocket Tester Kit includes the following items:

Apera Instruments PH60 pH Tester Kit contents laid out in its carrying case

Image: The complete PH60 pH Tester Kit, showing the meter, calibration solutions, bottles, and accessories neatly organized in its hard-shell case.

3. Setélan

3.1 Pamasangan batré

The PH60 tester requires 4 AAA batteries (included). To install or replace batteries:

  1. Unscrew the battery cap at the top of the meter.
  2. Pasangkeun 4 batré AAA, pastikeun polaritasna bener (+/-).
  3. Pasang deui tutup batréna kalayan pageuh.
PH60 pH tester with battery compartment open, showing AAA batteries

Image: The top section of the PH60 pH tester with the cap removed, revealing the battery compartment for AAA batteries.

3.2 Initial Probe Preparation

Before first use or if the tester has not been used for a long time (over 1 month):

  1. Pull out the battery insulation slip (if present) and remove the probe cap.
  2. Rinse the probe thoroughly with pure water (preferably distilled or deionized water, RO water is acceptable). Shake off excess water.
  3. If the tester hasn't been used for a long time (over 1 month), soak the probe in the 3M KCL soaking solution (included in the kit) for 15 minutes.
  4. Proceed to calibrate the meter before testing.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This video demonstrates the initial preparation steps, including removing the battery insulation slip, rinsing the probe, and soaking it in KCL solution if necessary (0:13 - 0:55).

4. Parentah Operasi

4.1 pH Calibration

The PH60 tester features 1-3 point automatic calibration with auto-buffer recognition. Always calibrate pH 7.00 first. If you need to calibrate pH 4.00 or pH 10.01, do so immediately after pH 7.00 calibration. Do NOT turn the meter off between pH 7.00 and pH 4.00/10.01 calibrations.

  1. Pour pH buffer solutions (pH 4.00, pH 7.00, pH 10.01) into the corresponding calibration vials to about half volume.
  2. Bilas usik dina cai murni tur ngocok kaluar kaleuwihan cai.
  3. Short press the power button to turn on the tester.
  4. Dip the probe into the pH 7.00 buffer solution. Make a quick stir, then let it stand.
  5. Long press the 'CAL' button to enter calibration mode. The screen will turn green.
  6. Wait for the reading to stabilize (a smiley icon will appear on the screen), then short press 'CAL' to finish the first point calibration. The tester will return to measurement mode, and an 'M' icon will appear at the bottom left, indicating a successful first-point calibration.
  7. Without turning off the tester, rinse the probe and repeat steps 4-6 for pH 4.00 buffer solution. An 'L' icon will appear for successful 2-point calibration.
  8. Without turning off the tester, rinse the probe and repeat steps 4-6 for pH 10.01 buffer solution. An 'H' icon will appear for successful 3-point calibration.
Step-by-step guide on how to calibrate the PH60 pH tester

Image: Detailed infographic illustrating the 3-point calibration process for the PH60 pH tester.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates the pH calibration process (0:56 - 4:26).

4.2 Pangukuran pH

To take a pH measurement:

  1. Short press the power button to turn on the tester.
  2. Bilas usik dina cai murni tur ngocok kaluar kaleuwihan cai.
  3. Insert the probe into your sample solution, make a quick stir, and hold still.
  4. Take the reading as the measurement when it is fully stabilized (a smiley icon appears and stays on screen).
  5. If the Auto-Hold function is turned on (in parameter setting P4), the reading will be automatically locked when it's stable for more than 10 seconds. The 'HOLD' icon will stay on. Short press the 'CAL' button to cancel Auto-Hold and continue measuring.
  6. Short press the 'MODE' button to switch to mV mode, which is for ORP testing or technical analysis (an ORP probe is needed for ORP measurements).
Hand holding the PH60 pH tester, showing a pH reading of 7.00

Image: A hand holding the PH60 pH tester, displaying a pH reading of 7.00 and a temperature of 25.0°C.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates the pH measurement process (4:26 - 5:33).

5. Pangropéa

5.1 Pembersihan Probe

The accuracy of the tester depends on the cleanliness of the probe. Always thoroughly rinse off the probe before and after each measurement with pure water.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates probe cleaning procedures (5:33 - 6:36).

5.2 Panyimpenan Probe

Proper storage is crucial for maintaining probe longevity and accuracy.

IMPORTANT: NEVER store the probe in pure water (like tap, RO, distilled, or deionized water) as they could damage the pH sensor. If this happens, immediately soak the pH probe in the 3M KCL soaking solution overnight to restore. Then re-calibrate it before using. Pure water is only for rinsing off the probe, NOT for storage.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates proper probe storage (6:36 - 7:29).

6. Cara ngungkulan

Ieu sababaraha masalah umum sareng solusina:

7. Spésifikasi

FiturNilai
Dimensi produk7 x 1.6 x 1.6 inci
Item Beurat10 ons
Nomer Modél ItemAI311
Batré4 batré AAA diperlukeun (kaasup)
Akurasi pH± 0.01 pH
Rentang pH-2.00 ka 16.00 pH
Rentang Suhu32 nepi ka 122 ° F (0 nepi ka 50 ° C)
Otomatis Daya-Pareuman8 menit tanpa operasi

8. Garansi jeung Rojongan

Apera Instruments is committed to providing reliable products and ongoing support to ensure you achieve the most accurate results. For warranty information, technical assistance, or any product-related inquiries, please refer to the official Apera Instruments website or contact their customer support directly. Contact details are typically provided on the product packaging or within the full user manual.

Dokumén Patali - AI311

Praview Apera PH60 pH Tester User Manual - Guide and Specifications
Comprehensive user manual for the Apera Instruments PH60 Premium pH Tester, covering setup, calibration, measurement, troubleshooting, and specifications.
Praview Apera EC20 Pocket Conductivity Tester: Instruksi Manual & Guide
Instruksi manual komprehensif pikeun Apera EC20 Pocket Conductivity Tester. Diajar ngeunaan setelan, kalibrasi, pangukuran konduktivitas, spésifikasi téknis, sareng inpormasi jaminan.
Praview Apera Instruments PH20 Value pH Tester Kit Manual Pamaké
Manual pamaké komprehensif pikeun Apera Instruments PH20 Value pH Tester Kit, ngawengku pamasangan batré, fungsi keypad, calibration, pangukuran, setélan parameter, ngungkulan masalah, spésifikasi teknis, sarta informasi jaminan.
Praview Apera Instruments 5052 Seawater Salinity Pocket Tester Operation Manual
Detailed operation manual for the Apera Instruments 5052 Seawater Salinity Pocket Tester, covering technical parameters, calibration, measurement procedures, important notes, and warranty information.
Praview Apera PH60-HT pH Meter User Manual: High Temperature & Caustic Solutions
Comprehensive user manual for the Apera PH60-HT pH Meter, designed for accurate pH measurement in high temperature (up to 140°C) and caustic solutions. Covers kit contents, operation, calibration, electrode care, parameter settings, technical specifications, troubleshooting, and warranty.
Praview Apera Instruments PH60 Premium pH Tester User Manual
Comprehensive user manual for the Apera Instruments PH60 Premium pH Tester, covering setup, calibration, measurement, maintenance, troubleshooting, and technical specifications.