Apera Instruments AI3115

Apera Instruments 201DJ-C AI3115 Double-Junction pH Electrode User Manual

1. Bubuka

This manual provides essential instructions for the proper use, maintenance, and troubleshooting of the Apera Instruments 201DJ-C AI3115 Double-Junction pH Electrode. Please read this manual thoroughly before using the electrode to ensure accurate measurements and extend its lifespan.

The Apera Instruments 201DJ-C AI3115 Double-Junction pH Electrode is specifically designed for long-term pH testing in challenging environments. Its robust construction makes it suitable for applications involving turbid water, wastewater, pools, spas, and general environmental monitoring. The double-junction structure is engineered to minimize clogging and contamination, which are common issues in such samples.

Pertimbangan penting: This electrode is not recommended for measurements in solutions with very low ion concentrations (e.g., deionized or distilled water), highly alkaline solutions (pH > 12), strong corrosive liquids, or for prolonged measurements in high-temperature solutions (above 60°C).

The electrode is compatible with all pH meters that utilize BNC connectors. Please note that a temperature sensor is not integrated into this electrode. For accurate automatic temperature compensation (ATC), a separate temperature sensor is required, or the temperature must be manually entered into your pH meter.

2. Produk Langkungview

The Apera Instruments 201DJ-C AI3115 is a high-performance combined pH electrode featuring a double-junction design for enhanced stability and accuracy in complex samples.

Apera Instruments 201DJ-C AI3115 Double-Junction pH Electrode

Figure 1: Apera Instruments 201DJ-C AI3115 Double-Junction pH Electrode. This image shows the full electrode with its blue body and protective storage cap.

Fitur konci:

  • Rentang pangukuran: 0-14 pH, 0-80 °C
  • Bahan perumahan: Polycarbonate for durability
  • diafragma: Double/Ceramic for reduced clogging
  • éléktroda rujukan: Ag/AgCl for stable potential
  • Panyambung: BNC, ensuring wide compatibility
Features of 201DJ-C Double-Junction pH Electrode

Figure 2: Key features of the 201DJ-C Double-Junction pH Electrode, highlighting its durable, accurate, stable, and fast response characteristics.

Electrode Design:

The double-junction design provides an additional barrier between the sample and the reference electrolyte, significantly reducing the risk of contamination and prolonging the electrode's life, especially in samples containing proteins, sulfides, or heavy metals.

pH Sensor Shield

angka 3: Tutup-up view of the pH sensor shield, designed to protect the sensitive glass bulb.

The electrode comes with a storage bottle containing a specialized solution to keep the sensor hydrated and in optimal condition when not in use.

Electrode with storage bottle

Figure 4: The electrode is supplied with a storage bottle, ensuring the sensor remains in optimal status.

The BNC connector ensures broad compatibility with various pH meters.

Panyambung BNC

Figure 5: The BNC connector, compatible with all BNC pH meters, provides a secure and standard connection.

3. Setélan

  1. Ngabongkar bungkusan: Carefully remove the electrode from its packaging. Inspect for any visible damage.
  2. Persiapan: Remove the storage bottle from the electrode tip by twisting and pulling gently. Rinse the electrode tip with distilled or deionized water. Do not wipe the glass bulb, as this can create static electricity and affect readings.
  3. Sambungan: Sambungkeun konektor BNC éléktroda kana input BNC anu saluyu dina méter pH anjeun. Pastikeun sambunganna aman.
  4. Temperature Sensor (if applicable): If your pH meter supports automatic temperature compensation (ATC) and you have a separate temperature sensor, connect it to your meter. Otherwise, you will need to measure the sample temperature manually and input it into the pH meter.
  5. Conditioning: For optimal performance, immerse the electrode in a pH 7.00 buffer solution for at least 30 minutes before initial use or after prolonged storage. This helps to rehydrate the glass membrane.
  6. Kalibrasi: Calibrate your pH meter with the electrode using at least two standard buffer solutions (e.g., pH 7.00 and pH 4.00 or pH 10.00). Refer to your pH meter's instruction manual for specific calibration procedures.

4. Parentah Operasi

  1. Bilas: Before each measurement, rinse the electrode thoroughly with distilled or deionized water to remove any residual buffer or sample.
  2. beuleum: Celupkeun ujung éléktroda kana sample solution, ensuring that the glass bulb and the junction are fully submerged. Stir the solution gently to ensure homogeneity and allow the reading to stabilize.
  3. Baca: Once the reading on your pH meter stabilizes (usually indicated by a stable icon or a consistent reading), record the pH value.
  4. Kompensasi Suhu: If using ATC, ensure the temperature sensor is also immersed in the sample. If manual temperature compensation is used, measure the sample temperature and input it into the pH meter before taking the pH reading.
  5. Bilas saatos dianggo: Saatos unggal pangukuran, bilas éléktroda sacara saksama nganggo cai sulingan atanapi cai deionisasi.
  6. Panyimpenan: Place the electrode back into its storage bottle containing the appropriate storage solution. Ensure the tip is fully immersed.

5. Pangropéa

5.1 Cleaning the Electrode

  • Bersihan rutin: Bilas éléktroda ku cai sulingan atanapi cai deionisasi saatos unggal panggunaan.
  • General Contamination: For general dirt or deposits, soak the electrode in a warm (not hot) solution of mild detergent for 15-30 minutes, then rinse thoroughly with distilled water.
  • Oil/Grease: Soak the electrode in ethanol or methanol for 15-30 minutes, then rinse thoroughly with distilled water.
  • Deposit protéin: Soak the electrode in a 0.1% pepsin solution in 0.1M HCl for 15-30 minutes, then rinse thoroughly with distilled water.
  • Deposit Anorganik: Soak the electrode in 0.1M HCl or 0.1M HNO3 for 15-30 minutes, then rinse thoroughly with distilled water.

After cleaning, recondition the electrode in pH 7.00 buffer for at least 30 minutes before recalibrating.

5.2 Panyimpenan

Always store the electrode with its tip immersed in the original storage solution (3M KCl solution or pH 4 buffer with KCl). Never store the electrode dry or in distilled/deionized water, as this will dehydrate the glass membrane and shorten its lifespan.

5.3 Electrode Lifespan

The lifespan of a pH electrode is typically 1-2 years, even with proper maintenance. Factors such as frequent use, harsh samples, and high temperatures can shorten its life. Replace the electrode when it can no longer be calibrated accurately or provides unstable readings after proper cleaning and conditioning.

6. Cara ngungkulan

MasalahMungkin Cukang lantarananaSolusi
Slow response or unstable readingsDehydrated glass membrane; Clogged junction; Contaminated buffer; Air bubbles in electrodeRecondition electrode in pH 7 buffer; Clean electrode; Replace buffer; Gently tap electrode to remove bubbles
Bacaan teu akuratImproper calibration; Contaminated buffer; Electrode aging; Temperature not compensatedRecalibrate with fresh buffers; Replace buffers; Replace electrode; Ensure proper temperature compensation
No reading or 'Error' messageLoose connection; Broken electrode; Meter malfunctionCheck BNC connection; Inspect electrode for damage; Test with another electrode or contact meter manufacturer
Mapay-mapay bacaanInsufficient conditioning; Contaminated junction; Temperature fluctuationsAllow more time for conditioning; Clean electrode; Ensure stable sampsuhu na

7. Spésifikasi

ParameterNilai
Nomer modélAI3115
Rentang pHpH 0 - 14
Rentang Suhu0 - 80 °C
Bahan PerumahanPolikarbonat
DiafragmaDouble / Ceramic
Éléktroda RujukanAg/AgCl
Tipe PanyambungBNC
Ukuran (L x W x H)16 x 1.27 x 1.27 cm
Beurat109 gram
ProdusénShanghai Sanxin Instrumentation
Negara asalCina

8. Garansi jeung Rojongan

For warranty information and technical support, please refer to the documentation provided with your specific purchase or visit the official Apera Instruments website. General support and contact information can typically be found on the manufacturer's product pages.

Dokumén Patali - AI3115

Praview Apera LabSen 241-180 pH Electrode User Manual
User manual for the Apera Instruments LabSen 241-180 Micro pH Electrode, detailing its features, technical specifications, usage, maintenance, and limited warranty.
Praview LabSen 373 Flat pH Electrode User Manual | Apera Instruments
User manual for the Apera Instruments LabSen 373 Flat pH Electrode, detailing technical specifications, usage, maintenance, and warranty information for accurate surface pH measurements.
Praview Apera Instruments 2310-C High-Range Conductivity Electrode User Manual
User manual for the Apera Instruments 2310-C High-Range Conductivity Electrode, detailing its features, technical specifications, installation, usage, maintenance, and limited warranty.
Praview Apera LabSen 871 pH Electrode User Manual for Non-Aqueous Solutions
User manual for the Apera LabSen 871 pH Electrode, detailing its features, specifications, preparation, measurement procedures, cleaning, maintenance, and warranty for non-aqueous solutions and organic solvents.
Praview Apera LabSen 241-6 Semi-Micro pH Electrode User Manual
User manual for the Apera LabSen 241-6 Semi-Micro pH Electrode, detailing its features, technical specifications, usage, maintenance, and warranty information for micro sampanalisis na.
Praview Apera LabSen 845 pH/Temp. Electrode User Manual
User manual for the Apera LabSen 845 pH/Temp. Electrode, detailing its features, technical specifications, usage instructions, cleaning procedures, maintenance, and warranty information for high-precision pH measurements.