APEX Domus 4 (QA-00050/D4)

Buku Panduan Pangguna Kasur APEX Domus 4 Anti-Decubitus

Model: Domus 4 (QA-00050/D4)

1. Bubuka

The APEX Domus 4 Anti-Decubitus Mattress with Compressor is designed to provide effective pressure redistribution and relief for patients at high risk of developing pressure ulcers, or for the treatment of existing pressure ulcers up to Grade IV. This system aims to improve patient comfort and quality of life by preventing and managing skin breakdown associated with prolonged bed rest.

This manual provides essential information for the proper setup, operation, maintenance, and troubleshooting of your Domus 4 system. Please read it thoroughly before use to ensure safe and effective operation.

2. Émbaran Kasalametan penting

Always follow basic safety precautions when using electrical products, especially when children are present. To reduce the risk of electrocution, burns, fire, or injury to persons:

  • Read all instructions before using the mattress system.
  • Do not operate the compressor if it has a damaged cord or plug, if it is not working properly, or if it has been dropped or damaged.
  • Tetep ari jauh ti surfaces dipanaskeun.
  • Ulah ngahalangan liang hawa atawa nempatkeun dina permukaan anu lemes dimana liangna bisa kahalang.
  • Entong nganggo di luar atanapi beroperasi dimana produk aerosol (semprotan) dianggo atanapi dimana oksigén dikaluarkeun.
  • Ensure the mattress is placed on a stable, flat bed frame.
  • Regularly inspect the mattress and compressor for any signs of wear or damage.
  • Consult a healthcare professional for appropriate patient positioning and care.

3. Komponén Sistem

The APEX Domus 4 system consists of the following main components:

  • Alternating Pressure Mattress: Composed of individual air cells that inflate and deflate in an alternating pattern to redistribute pressure.
  • Unit Kompresor: The pump that inflates and deflates the air cells, controlling the pressure and cycle time.
  • Antimicrobial Cover: A protective, vapor-permeable, water-resistant, and flame-retardant cover for the mattress.
  • Selang hawa: Connect the compressor to the mattress.
  • CPR Valve: A quick-release valve for rapid deflation in emergency situations.
  • Angkutan Plug: Used to maintain air in the mattress during patient transport or power outages.
APEX Domus 4 Anti-Decubitus Mattress System

Gambar 3.1: Leuwihview of the APEX Domus 4 Anti-Decubitus Mattress System, showing the mattress on a bed frame with the compressor unit attached.

APEX Domus 4 Compressor Unit

Figure 3.2: The APEX Domus 4 compressor unit, featuring control panel and power input.

APEX Domus 4 Mattress Air Cells

Figure 3.3: The APEX Domus 4 mattress, showcasing its individual air cells designed for pressure redistribution.

CPR Valve on Mattress

Figure 3.4: Close-up of the red CPR valve on the mattress, used for rapid deflation in emergencies.

4. Parentah Setup

Follow these steps to set up your APEX Domus 4 Anti-Decubitus Mattress system:

  1. Nyiapkeun pigura ranjang: Place the mattress on a standard bed frame. Ensure the bed frame is clean and free of sharp objects that could damage the mattress.
  2. Buka gulungan kasur: Carefully unroll the mattress on the bed frame with the air cell side facing up. Ensure the air hoses are positioned towards the foot end of the bed.
  3. Secure the Mattress: Use the adjustment straps located on the underside of the mattress to secure it to the bed frame, preventing slippage.
  4. Connect the Compressor: Hang the compressor unit securely at the foot end of the bed using its integrated hangers.
  5. Sambungkeun Selang Udara: Connect the air hoses from the mattress to the air outlets on the compressor unit. Ensure a secure and tight connection.
    Connecting Mattress to Compressor

    Figure 4.1: Demonstrates connecting the air hoses from the mattress to the compressor unit.

  6. Plug in the Compressor: Plug the compressor's power cord into a grounded electrical outlet (AC 220-240V, 50 Hz).
  7. Inflasi Mimiti: Turn on the compressor using the On/Off button. The mattress will begin to inflate. Allow approximately 20-30 minutes for the mattress to fully inflate and stabilize before placing a patient on it.

5. Parentah Operasi

5.1. Control Panel Leuwihview

The compressor unit features a control panel for adjusting settings and monitoring system status. Refer to the labels on your specific unit for exact button functions.

5.2. adjustment tekanan

Adjust the pressure setting based on the patient's weight. The compressor allows for pressure range adjustment for patients weighing between 40 kg and 200 kg. Use the pressure adjustment dial or buttons on the compressor to select the appropriate setting for optimal pressure redistribution.

5.3. Therapy Modes

The Domus 4 offers three therapy modes:

  • Alternating Pressure Mode: This is the primary mode, where air cells inflate and deflate in an alternating cycle (typically 10 minutes) to continuously redistribute pressure and stimulate blood flow.
  • Modeu statik: In this mode, all air cells remain inflated to a constant pressure, providing a firm, stable surface. This mode is useful for patient transfers or when a stable surface is temporarily required.
  • Anti-Shearing Mode: This mode helps reduce shearing forces on the patient's skin, often by slightly reducing pressure in certain areas or adjusting cell inflation patterns.

5.4. Fungsi Husus

  • Max Firm: This function inflates all cells to maximum firmness, useful for patient care activities like dressing changes or transfers.
  • Konci Panel: Activates a lock function on the control panel to prevent accidental changes to settings.
  • Modeu Angkutan: When the compressor is disconnected, the mattress can maintain air for a period using the transport plug. Insert the transport plug into the air hose connection point on the mattress to seal the air within the cells.
    Transport Plug for Mattress

    Figure 5.1: The transport plug, used to seal air within the mattress cells for patient transport or power outages.

  • Fungsi CPR: In an emergency requiring CPR, pull the red CPR valve to rapidly deflate the mattress, providing a firm surface for resuscitation.
Patient's foot on mattress showing pressure relief

Figure 5.2: Illustrates the pressure redistribution effect of the mattress on a patient's foot, highlighting its therapeutic efficacy.

6. Pangropéa sarta perawatan

6.1. Cleaning the Mattress Cover

The mattress cover is designed for easy cleaning. It can be machine washed. Follow these guidelines:

  • Remove the cover from the mattress.
  • Machine wash according to the care label instructions (typically gentle cycle with mild detergent).
  • Entong dipasihan pamutih.
  • Air dry or tumble dry on low heat. Ensure the cover is completely dry before reattaching.

6.2. Cleaning the Compressor Unit

Wipe the exterior of the compressor unit with a damp cloth and mild disinfectant. Do not immerse the unit in water or spray liquids directly onto it.

6.3. Panyimpenan

Upami nyimpen sistem kanggo waktos anu lami:

  • Kempis kasur lengkep.
  • Ngabersihan sadaya komponén sacara saksama.
  • Fold or roll the mattress neatly.
  • Store the mattress and compressor in a cool, dry place, away from direct sunlight and extreme temperatures.

7. Cara ngungkulan

This section provides solutions to common issues you might encounter with your Domus 4 system.

MasalahMungkin Cukang lantarananaSolusi
Mattress not inflating or losing air.Air hoses not securely connected.
Compressor not powered on.
Leak in mattress or hoses.
CPR valve open.
Ensure hoses are tightly connected.
Check power cord and On/Off switch.
Inspect mattress and hoses for punctures; repair if possible or contact support.
Ensure CPR valve is closed.
Compressor making unusual noise.Compressor unit not stable.
Gangguan internal.
Ensure compressor is hung securely and not vibrating against a surface.
Upami aya sora anu terus-terusan, eureunkeun panggunaan sareng hubungi layanan palanggan.
Alarms sounding (Low Pressure, Power Failure, System Fault).Low pressure: Leak or patient weight exceeding capacity.
Power failure: Power disconnected.
System fault: Internal error.
Check for leaks, adjust pressure setting. Ensure power cord is connected. For system fault, restart unit; if alarm persists, contact support.
Mattress feels too firm or too soft.Incorrect pressure setting for patient weight.Adjust the pressure setting on the compressor unit according to the patient's weight (40-200 kg range).

If you encounter a problem not listed here, or if the issue persists after attempting the suggested solutions, please contact customer support.

8. Spésifikasi Téknis

8.1. Mattress Specifications

  • Ukuran (L x W x H): 200 x 85 x 20.32 cm
  • beurat: 6.9 kg
  • Sél: 20 independent 20.32 cm Nylon PU cells
  • Static Head Zone: 4 sél
  • Ventilated Cells: 9 sél
  • Bahan dasar: Nylon PVC
  • Fungsi CPR: Sumuhun
  • Pituduh Kabel: Sumuhun
  • Adjustment Straps for Bed Frame: Sumuhun
  • Internal Safety Springs: Sumuhun
  • External Protection Tube: Sumuhun
  • Flame Retardant Certificates: EN 597-1 jeung EN 597-2

8.2. Compressor Specifications

  • Ukuran (L x W x H): 29 x 18.5 x 12.6 cm
  • beurat: 2.2 kg
  • Voltage: AC 220-240V, 50 Hz
  • Aliran Udara Max: 8 liter/menit
  • Pressure Range Adjustment: 40-200 kg patient weight
  • Alternating Cycle Time: 10 menit
  • Terapi: Alternating, Static, Anti-shearing
  • fungsi: Max Firm, Panel Lock
  • Alarm: Low Pressure, Power Failure, System Fault
  • Tingkat Noise: 26 dBA
  • Tombol Hurung/Pareum: Sumuhun
  • Modeu Angkutan: Sumuhun
  • Gantungan: Sumuhun
  • Casing: Seuneu-retardant ABS
  • Sertipikat: Class II and Type BF electrical shock protection, IP21 electrical safety standards

8.3. Cover Specifications

  • Bahan: Padded Nylon PU with ionic silver (Ag+) antibacterial treatment
  • Panutup: Zipper with flap
  • Tahan cai: Sumuhun
  • Vapor Permeable: Sumuhun
  • Kontrol Inféksi: Anti-mold and anti-fungal
  • Biokompatibilitas: Sumuhun
  • Parentah perawatan: Nyeuseuh mesin

9. Garansi jeung Rojongan

9.1. Émbaran jaminan

The APEX Domus 4 Anti-Decubitus Mattress system, including both the mattress and the compressor, comes with a garansi 2 taun from the date of purchase. This warranty covers manufacturing defects under normal use. It does not cover damage resulting from misuse, accidents, unauthorized repairs, or failure to follow the instructions in this manual.

9.2. Rojongan Palanggan

For technical assistance, warranty claims, or to inquire about replacement parts, please contact your authorized dealer or the manufacturer's customer service. Please have your product model number (Domus 4 or QA-00050/D4) and purchase date available when contacting support.

Manufacturer: Apex

Brand: Queraltó

Dokumén Patali - Domus 4 (QA-00050/D4)

Praview APEX XT Auto Provider Mode Quick Guide
A quick guide for medical professionals on how to use the APEX XT Auto device in Provider Mode, including setup navigation and resetting compliance meters.
Praview Buku Panduan Korsi Pijat APEX-V2 - Panduan Persiapan, Operasi, sareng Pangropéa
Comprehensive instruction manual for the APEX-V2 massage chair, covering safety precautions, installation, operation, settings, specifications, troubleshooting, cleaning, and maintenance. Learn how to set up and use your APEX-V2 massage chair effectively.
Praview Buku Panduan Sistem CPAP Otomatis APEX XT
Manual instruksi lengkep pikeun Sistem APEX XT Auto CPAP (Modél 9S-005720), anu ngajelaskeun sacara rinci setelan, operasi, pangropéa, ngungkulan masalah, sareng spésifikasi téknis pikeun ngubaran Obstructive Sleep Apnea (OSA).
Praview Buku Panduan Pangguna Pompa APEX VAC: Pitunjuk sareng Panduan Kasalametan
Buku panduan pangguna anu lengkep pikeun Pompa APEX VAC, anu ngajelaskeun operasi, kaamanan, spésifikasi téknis, sareng cara ngungkulan masalah pikeun modél VacPro, VacMaxi, sareng VacPlus ti Apex Medical Corp.
Praview Manual Pamilik Sistem Kaamanan DESTINY
Buku panduan ieu pikeun Sistem Kaamanan Apex DESTINY nyayogikeun pitunjuk anu lengkep ngeunaan pamasangan, operasi, modeu arming (AWAY, HOME, NIGHT), aksés jauh, fungsi bypass, sareng fitur otomatisasi bumi pilihan.
Praview Manual Operasi Pamaké Kursi Pijat Apex AP-Pomp
Manual operasi pamaké komprehensif pikeun korsi urut Apex AP-Pomp, ngawengku precautions kaamanan, pangropéa, instalasi, fungsi controller, ngungkulan, sarta spésifikasi.